How to operate an ventilator machine?

1 connecting power, oxygen piping, compressed air piping (or detection of oxygen pressure)
2.Join gas pipeline, a heated humidifier add water
3 set breathing machine commonly ventilation parameters
Breathing pattern: control or auxiliary (A / C or SIMV)
Tidal volume: 350 ~ 600ml, respiratory rate 16 to 20 times / min
Respiratory ratio: 1:1.5 to 2.5, the oxygen concentration of 30% to 50% (or set the minute ventilation,
Inspiratory time), PEEP (PEEP) or pressure support (PSV)
4 Set the ventilator monitoring and alarm parameters
5.Open the ventilator power switch, observe the ventilator working condition
(6) gas pipeline connection simulated lung, the observed catheter balloon inflation and deflation, check the gas pipeline is smooth, with or without leakage, observe whether the ventilation parameters match the settings
7.Clear the airway secretions, adjust the fixed ventilation mask, inflatable airbag endotracheal tube or tracheostomy tube
8 Removal of lung simulator: gas pipelines and ventilation mask, endotracheal intubation or casing connections, auscultation of breath sounds lungs are symmetrical to test artificial airway patency closed
9.Observed ventilator is working properly, the patient vital signs tend to be normal.