XFA6000Intelligent intelligent automatic blood cell analyzer)

◇Added : 2013-11-11 10:24:37  ◇hits : 3
Blood instrument used in the development of our country already has more than 50 years, the medical equipment in hospitals in our country has a very wide range of use, and the boot at positive pressure suction fault is perhaps blood instrument one of the most common problems, or other similar problems, plant below small make up to you in detail under the blood meter pressure problems common solutions:

With seal components, for example, store liquid bottle, waste liquid bottle, buffer bottle mixed pool etc. These demands because these all positive pressure and negative pressure drive, sealing ring and simple device does not reach the designated position may be aging, and many maintenance manual, we will notify the operator to clear often arrange the local, simple form device does not reach the designated position forming gas leakage. Of course, the leakage of cylinder is also nots allow to ignore.


(perlong medical - XFA6000Intelligent intelligent automatic blood cell analyzer)

After check the above parts, doubt no problem, will do further analysis and disposal, first listen to the first ring, blood instrument have gas leakage or inhaled sound, very different from these two kinds of sound, bad to describe in words, my feeling is "poof -" sound is leaked, "bursts -- -- -- -- -- -" sound is inhaled, in order to distinguish is the positive pressure is still a negative pressure, and select a size for troubleshooting.

See there are many ways to, usually choose block elimination process, namely in the section of the pipeline to see whether the fault, if the report is this a question, then, of course, a line drawing to understand, otherwise, the clamp is a main line, ha ha, how to quote. If a pressure gauge, the job becomes easier, first on the pressure gauge received positive pressure may be negative pressure interface, so as to distinguish whether lack of pump pressure, if not, then the disposal of pump first, the original pressure is normal, according to the line graph one by one, the branch node clamp.


评论

此博客中的热门博文

630mA Radiography X ray Machin | price of DR system PLD6000

microplate reader 9602

Medical Mammography X Ray System (BTX-9800 series)